No
|
ITEMS
|
QTY
|
SPECIFICATIONS
|
Brand/Make
|
|
1
|
Table top Anesthesia mobile Machine
|
1
|
Material: Stainless Steel, High Impact Plastic Application: Adult and Children Respiratory rate: 4~60bmp Power: 220V±10%,50Hz±1% Working Mode: For the Closed System, Semi-Closed System and Semi-Open System Oxygen Flush: 35L/min--75L/min Working Mode: Manual Control or Automatic Control Respiratory Modes: IPPV,SIGH(Deep Breaths/Per Time After Breathing 60 Times), SIMV/F/2, PEEP(+0.5KPa -- +2KPa), SIPPV Tidal Volume: 30~1500ml Power: 220V±10%,50Hz±1% Alarms Parameters: Upper and Lower Airway Pressure Alarm; Upper and Lower Tidal Volume Alarm; Lack of Oxygen Alarm; Apnea; Power Outages I/E Ratio: 3; 1-1:4 Adjustable 3: 1-1:4 Adjustable Vaporizer: Two High Precise Vaporizers, scope of Concentration Regulation: 0.2 -- 5%VOL
|
Drager Visat 120/Dre medical
|
|
2
|
Laryngoscopes (Adult and Pediatric)
|
1
|
Includes a cylindrical handle, slightly ribbed, single piece, Ø > 25 mm The handle allows fitting of interchangeable blades of different sizes. The handle contains a robust on/off switch. Handle is made from non-ferrous material and sealed against ingress of liquids. Equipped with LED technology light source. Uses fibre optic for transmission of the light to the tip of the blade. Stud contact, fitting various sizes and types of depressors, compliant with ISO 7376. Blades are able to withstand autoclaving. With a set of four blades made from stainless steel: - MacIntosh type: Curved Nr 2, length 110mm ± 5 mm. - MacIntosh type: Curved Nr 3, length 135mm ± 5 mm. - MacIntosh type: Curved Nr 4, length 155mm ± 5 mm. - Miller type: Straight Nr 1, length 100mm ± 5 mm Designed for frequent and easy disassembly and disinfection with hospital-grade products
|
C Mac
|
|
3
|
Defibrilator with battery
|
1
|
- Should be a Low Energy Biphasic defibrillator monitor with Recorder, having capability to deliver shocks from 2 Joules to 200 Joules.
- Should monitor ECG through paddles, pads and monitoring electrodes and Defibrillate through pads and paddles.
- Should compensate for body impedance for a range of 25 to 150 ohms
- Should have a built in 50 mm strip printer
- Should have charging time of less than 5 seconds for maximum energy.
- Should have High resolution more than 8 inch Colour display for viewing monitoring parameters
like ECG, SpO2, NIBP and etCO2 with 4 waveform capability of 4 seconds.
- Both Adult and pediatric paddles should be available.
- Should have facility for self test/check before usage and set up function.
- Should have facility to monitor parameters like SpO2, NIBP and etCO2 along with non-invasive pacing (Demand & Fixed mode) facility.
- Should be able to upgrade the defibrillator for 12 lead ECG monitoring and ECG transmission.
- Power input to be 120-240VAC, 50-60 Hz Should have a battery capable of usage for at least five hours.
- Should have a battery capable of usage for at least 5 hours of monitoring.
|
Comen S8
|
|
4
|
Mobile vital signs machine (Monitor for patients)
|
2
|
MUST HAVE 6 Para: ECG, RESP, NIBP, SPO2, 2_TEMP, PR/HR Type Arrhythmic Analysis, Multileader ECG Waveforms Display in Phase, Real-time S_T segment analysis, pacemaker detection Drug calculation, and titration table; Efficient resistance to interference of defibrillator and electrosurgical cautery; SPO2 can test for 0.1% Weak; RA-LL impedance Respiration; Trend Coexist Display; OxyCRG Dynamic View Display; Bed to Bed view Display; Networking capacity and nurse calling system; Options of Central Monitoring System; Built-in rechargeable battery 2200mAh; 8" high-resolution color TFT LCD display; The large volume of tabular and graphic trends information storage and easy to recall; Anti-ESU, anti- defibrillator; Capture dynamiStandardorms. Standard Packing List 5-lead ECG cable (including RESP) Adult and pediatric finger SpO2 sensor Adult, pediatric, and baby NIBP cuff NIBP extension tube TEMP sensor AC power adaptor ECG electrodes Operation instruction must be with Patient Monitor Roll Stand With Basket
|
Mindray
|
|
5
|
Suction machine
|
2
|
Portable suction unit: Housing: Plastic molded cabinet Capacity: -700 mm Hg ± 10 mm regulable, flutter-free vacuum control knob, 35 Ltrs/min Pump Type: Double diaphragm pump Jars: 2 x 2.5 ltr. wide-mouthed (Polycarbonate) with a changeover lever. Provided with handle for hygiene while emptying/cleaning. Safety jar with mechanical / electronic overflow motor shut-off system and bacterial filter. Tubing: 10 mm ID x 2 meters. (PVC) Vacuum Gauge: Bourden type 6.25 cm dia 0-760 mm, Hg calibration. Power: 220 V, 50 Hz, Single Phase Noise Level: 60 dB A±3 Dimension & Weight: 46 x 32 x 86 cms, 23 Kg.
|
Accusure
|
|
6
|
Tourniquet (Adult and Pediatric)
|
1
|
Materials: Natural rubber. Strap: semi-rigid, elastic and tubular. Size: length 50cm, width 2cm, wall thickness 0.8-1mm
|
Ysenmed
|
|
7
|
Electro Surgical machine
|
1
|
Output Power: 150 W
Operating frequency (monopolar): 330 kHz/430 kHz
Power Rating: 500 VA ± 10 %
Monopolar cut operation (pure/ blend cut): 0 to 150 W (300 Ω)
Monopolar coagulation operation (forced): 0 to 80 W ( 500 Ω)
Monopolar coagulation operation (soft) 0 to 40 W (500 Ω)
Operating frequency (bipolar)
Bipolar coagulation
Biopolar
Monopolar cut (load 500 Ω)
Monopolar coagulation (load 500 Ω)
Bipolar (coagulation/AUTO) (load 100 Ω)
|
Surtron 200
|
|
8
|
Autoclave
|
1
|
23L
Power: 220VAC 50/60Hz . 4 kw
Type: table top
Temperature Range: 121-134 degrees
Camber capacity:20-25L
Max pressure: 0.22mpa
Timer: max 60minutes
Light indicator: Yes
Lack of water alarms: yes
Outer Dimensions: apprx. LWH 580mm*480mm*384mm
Chamber dimensions: apprx.250mm* 520mm
Temperature stability: <=1- degrees
Net weight: not more than 50kg.
Drainage: yes
Preset : yes.
|
Tuttnaeur
|